Baxter 450, Baxter SPS550, Baxter 1550 SPECIFICATIONS: Dimensions (HxWxL) 120cm x 5.8cm x 67.9cm Weight 81kg (178lbs.) Floor space 3.75 sq ft ENVIRONMENT Temperature 10(c) 50(F) to 35(c) 95(F) Relative humidity 30% to 95% ELECTRICAL SERVICE The Baxter machines are designed to meet all UL and CSA requirements. - Current requirements SPS550 50/60 Hz, VAC + 10% at 13 amps maximum SPS550 50/60 Hz may be wired for: -100 VAC + 10% at 14.4 amps -200 VAC + 10% at 7.2 amps -220 VAC + 10% at 6.5 amps -240 VAC + 10% at 6.0 amps - Electrical leakage Less than 50 microamperes at all voltages. - Dieletric strength 2500 VAC - Air conditioning load 1,785 BTU/hr or 2500 watts WATER SUPPLY REQUIREMENTS Water must meet AAMI standards for hemodialysis. - Supply pressure 35 kPa (5psi) to 700 Kpa (100 psi) - Internally regulated pressure - 70 Kpa (10psi) after the inlet water pressure regulator. - 100 KPa (15 psi) after the proportioning pump pressure regulator. |