Fresenius Products
Models Available:
Fresenius
2008-H, Fresenius 2008-E, Fresenius 2008-D, Fresenius 2008-C
- Product Description
When Fresenius designed the 2008 unit it was designed
to be a modular concept. This concept would allow any updates that
become available as technology advances. Once the upgrades are available,
users simply update the equipment. These upgrades allow patients the
latest in dialysis therapies.
- Features
- Modular component design.
- Ultra precise volumetric UF control.
- Blood flow rates of 0 to 600 ml/min.
- Variable dialysate flow rates.
- Proven reliability.
- Automatic blood pressure monitoring (optional).
- Sodium Variation up to 15%.
- Computer interface (optional).
- Specifications
- Dimensions (DxWxH) 50cm x 50cm x 114cm
- Weight 70 kg (155 lbs.)
- Voltage 110, 220vac 50/60 hz
- Current 12 amps max
- Power 1350 watt @110 vac
- Plug configuration NEMA 5-15 P
- Environmental
- Water pressure 1.0 to 7.0 bar
- Water temperature 10.0 c to 25 c
- Flow rate 900ml/min minimum
- Drain height 46cm (18 in.) off floor max
- Drain ID 4cm (1.5in.)
- Air conditioning load 900 BTU/hr.
- Performance
- Dialysate flow rate:
Minimum
300ml/min
Maximum
600ml/min
Standard 500ml/min
- Ultra filtration control:
W /
volumetric control 0-2000 ml/hr
W / TMP
control 4500 ml/hr
- Hot water rinse:
Temperature 85c
Display
range 80c to 100c
Flow
rate 300 ml/min
Time 20
min avg.
Venous
pressure range -50 to +250 mm/hg
Arterial pressure range -150 to +250 mm/hg
TMP
pressure range +60 to –540 mm/hg |